产品特性:测氧 | 是否进口:否 | 加工定制:否 |
品牌:hamilton | 型号:VisiTrace DO 225 | 订货号:243561 |
货号:VisiTrace DO 225 | 测量范围:/ | 测量精度:// |
分辨率:/// | 电源电压:220 | 用途:VisiTrace DO 225 |
是否跨境货源:否 |
Specifications
Specification Sheet
a-length | 120 mm |
---|---|
Electrical Connector | M12 |
Measurement Principle | Oxygen dependent luminescence quenching |
Measuring Range | 0 to 2 ppm (0 to 50 mbar) |
Accuracy at 25 °C | ± 0.5 ppb or 2 % whichever is greater |
Drift at Room Temperature | <1 % per week @ 100 ppb |
Sensor Cap | ODO Cap L0 |
Electrolyte | None |
Oxygen Consumption | None |
Temperature Sensor | NTC 22 kOhm |
Configurable Values | DO: mbar; %-sat; %-vol; ?g/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C, °F, K |
Diameter | 12 mm |
Process Connection | PG13,5 |
Wetted Parts | Stainless Steel 1.4435 EPDM (Ethylene propylene elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document |
Surface Quality of Steel | Ra<0.4 ?m (N5) |
4-20 mA Current Range | 3.5 to 22 mA |
Analog Interface 1 | Two wire sink 4-20 mA for DO, programmable |
4-20 mA Accuracy | <0.3 % + 0.05 mA |
Transmitting Power | Max. 4 dBm |
Digital Interface | Bluetooth 4.0 |
Operating Voltage | 18 to 30 VDC |
Serial Number | Yes |
Response Time t90% | <20 s in gas; <90 s in water |
Certificate | Yes, with measured current in oxygen |
ATEX Approval | No |
IECEx Approval | No |
Autoclavable | No |
CIP | Yes |
Steam Sterilizable | Yes, max. Temperature 140 °C |
Operating Temperature Range | -10 to 140 °C; the sensor provides no DO reading above 85 °C |
Pressure Range bar g | -1 to 12 bar |
Protection Rating | IP 68 |
Required Flow | None |
货号:243561
VisiTrace DO电极秉承了Hamilton光学溶氧电极响应速度快、容易维护等优点,摒弃了传统极谱式氧电极需要极化、测量受流速影响等缺陷,非常适合啤酒、电力等需要测量ppb级低氧的场所。
在啤酒厂麦汁发酵生产啤酒的过程中,酵母生长需要耗氧。但是在发酵末期啤酒氧化会导致啤酒的品味下降、货架期缩短。在啤酒厂,VisiTrace DO 电极的主要应用点是清酒过滤和灌装。为啤酒行业特殊设计的ODO Cap L0膜帽可以耐受常用消毒剂中的活性氯气和二氧化氯,在一些要求在CIP清洗后电极不能再拿出来校准的测量点,这个设计就非常有优势了。
在电厂,测量锅炉用水的含氧量可以***锅炉高效运转,节省运营费用。锅炉用水中即使很低的含氧量也会明显降低锅炉的整体寿命。如果锅炉用水中氯离子的含量高, 氧的腐蚀作用会更大。由此可见ppb级的低氧监控就非常重要了。
VisiTrace DO – 明智的选择
VisiTrace DO 电极通过内置的变送单元直接向控制系统输出***的测量信号。位于电极头部的内置变送单元存储了所有的电极信息,包括校准信息、自诊断信息等,使得校准与维护变得非常简单。
通过支持蓝牙 4.0 的无线设备(如iPhone等)可以查看电极的设置、校准、性能指标(耗损度)等信息。